CRCST Sterilization Process practice questions
Steam and low-temperature sterilization methods and cycles, sterilizer testing, loading and unloading, physical, chemical and biological monitoring, process failures, recalls and load documentation. This domain is about 21% of the exam. Try each question, then open the answer.
Read the Sterilization Process study guide
Question 1
- Wear heat-resistant gloves and pull the cart out by its handle
- Open the door fully and lean in to check the load while the steam clears
- Remove each pack by hand right away to check it for moisture
- Wear nitrile exam gloves and lift the hot trays straight off the shelves
Show answer
Answer: A. Wear heat-resistant gloves and pull the cart out by its handle
Loads, racks and chamber surfaces are very hot at the end of a cycle. Heat-resistant gloves and moving the load on its cart by the handle keep hands away from hot metal and steam.
Question 2
- Whether the incoming steam is at least 97% dry steam
- Whether wrapped packs are fully dry at the end of the cycle
- Whether air leaks into the chamber while it is under vacuum
- Whether bacterial spores are killed during the exposure phase
Show answer
Answer: C. Whether air leaks into the chamber while it is under vacuum
In a leak test the sterilizer pulls a vacuum, holds it, and measures how much the pressure rises. Air leaking in through seals, valves or piping can mix with the steam and keep it from contacting devices. The test is run as often as the manufacturer recommends (some call for daily, some weekly), and a failure means the sterilizer needs service.
Question 3
- Each day, in the last load of the day along with other packs
- Once a week, in a chamber that is fully loaded
- Daily, in an empty chamber, in the first cycle after warm-up
- Only after a biological indicator has come back positive
Show answer
Answer: C. Daily, in an empty chamber, in the first cycle after warm-up
The Bowie-Dick test (a dynamic air removal test) checks that air is removed and steam penetrates quickly. AAMI ST79 calls for it each day the sterilizer is used, in an empty chamber, before the first processed load. It runs after a warm-up cycle so the chamber is at temperature. A failed test takes the sterilizer out of use until the cause is fixed.
Question 4
- At least weekly, and preferably daily, per cycle type
- Only on days when the Bowie-Dick test has failed
- Only after the sterilizer has been repaired or moved
- Once a month, in the first load of each month
Show answer
Answer: A. At least weekly, and preferably daily, per cycle type
ST79 recommends routine BI monitoring at least weekly, and preferably every day the sterilizer is used, for each type of cycle run. Every load that contains an implant is also monitored with a BI PCD. BIs are the only monitors that directly show whether spores were killed.
Question 5
- The sterilizer failed, so all loads since the last BI must be recalled
- The test is invalid; it must be repeated and the cause investigated
- The control result does not matter if the processed BI is negative
- The load passed because both BIs show no growth
Show answer
Answer: B. The test is invalid; it must be repeated and the cause investigated
A control BI is expected to grow. If it does not, the spores, the media or the incubator may be faulty, so the negative result on the processed BI means nothing. The test is repeated with a good lot and a working incubator, and the event is documented and investigated.
Question 6
- Saturated steam that is at least 97% dry
- Steam mixed with some air so the heat spreads evenly
- Wet steam with a dryness of about 85%
- Superheated steam that is hotter than its pressure calls for
Show answer
Answer: A. Saturated steam that is at least 97% dry
Saturated steam holds the most heat it can at a given pressure without extra liquid water. Steam that is 97% or more dry releases its latent heat when it condenses on cooler surfaces, which kills microorganisms quickly. Wet, superheated or air-mixed steam all reduce effectiveness.
Question 7
- 4 minutes
- 10 minutes
- 15 minutes
- 30 minutes
Show answer
Answer: D. 30 minutes
At 250 F (121 C) in a gravity cycle, wrapped instruments typically need 30 minutes of exposure, plus drying time. The lower temperature and passive air removal mean a long exposure. Always follow the device and packaging IFU, which may require more.
Question 8
- To cool the items down after a high-temperature cycle
- To dry condensate that forms during steam exposure
- To remove EO residues that are toxic to patients and staff
- To wait until the biological indicator result is read
Show answer
Answer: C. To remove EO residues that are toxic to patients and staff
Plastics, rubber and other materials absorb EO during the cycle. Aeration passes warm, filtered air over the load so the gas leaves before anyone handles the items or they touch a patient. Aeration time and temperature come from the IFU, often about 12 hours at 50 C.
Question 9
- Polypropylene sterilization wrap
- A standard paper-plastic peel pouch
- A rigid container validated for the system
- A Tyvek-type (spunbonded polyolefin) pouch
Show answer
Answer: B. A standard paper-plastic peel pouch
Paper, linen and other cellulose materials absorb hydrogen peroxide, which can cancel the cycle or leave too little sterilant to reach the device. Packaging for these systems is non-cellulose, such as Tyvek-type pouches and polypropylene wrap. Always check both the sterilizer and packaging IFUs.
Question 10
- Wrapped and placed in sterile storage for later use
- Aerated for 12 hours before it is used on a patient
- Used right away, since it is wet and unpackaged
- Placed in a peel pouch and labeled with a load number
Show answer
Answer: C. Used right away, since it is wet and unpackaged
Liquid chemical sterilant systems process items unwrapped and finish with a rinse, so the device comes out wet. It cannot stay sterile in storage, so it is carried in the system's tray to the point of use and used immediately. This is often called just-in-time processing.
Question 11
- On the top shelf, near the back, close to the steam inlet
- On the cart outside the chamber, as the unexposed control
- Inside the largest instrument tray on the middle shelf
- On the bottom shelf, near the front, over the drain
Show answer
Answer: D. On the bottom shelf, near the front, over the drain
The area above the drain at the bottom front of the chamber is where air and condensate collect. That makes it the hardest spot for steam to reach. Placing the PCD there makes it the toughest challenge in the load.
Question 12
- Report it, take the incubator out of use, and repeat the BI tests
- Read the BIs at the usual time and record the results as normal
- Note the reading in the log and keep using it until its next PM
- Move the BIs to room temperature until the incubator is repaired
Show answer
Answer: A. Report it, take the incubator out of use, and repeat the BI tests
Incubator temperature is checked and recorded because the test organism grows only in a narrow range. Too cool an incubator can hide surviving spores, so a negative result from it cannot be trusted. The incubator is reported and removed from use, and the supervisor decides how to handle the affected loads. The BI tests are repeated in a working incubator.
Question 13
- Release it, since 10 minutes is the gravity exposure time at 270 F
- Release only the trays whose internal indicators changed color
- Hold the load, since it needed 15 minutes, and reprocess it
- Release it, since the printout shows the cycle completed
Show answer
Answer: C. Hold the load, since it needed 15 minutes, and reprocess it
Wrapped instruments in a gravity cycle at 270 F (132 C) typically need 15 minutes of exposure. A 10-minute exposure did not meet the required parameters, so the load is treated as non-sterile and reprocessed. The wrong cycle selection is reported so the cause can be corrected.
Question 14
- A Type 1 indicator on the outside of the tray, with verbal approval
- A PCD with a BI and Type 5 integrator, an exception form, and follow-up
- A passed Bowie-Dick from that morning, with the cycle printout attached
- The cycle printout alone, as long as the load has fully cooled
Show answer
Answer: B. A PCD with a BI and Type 5 integrator, an exception form, and follow-up
Early release of an implant is an exception, not routine. The load must be monitored with a PCD holding a BI and a Type 5 integrating indicator. The release is documented on an exception form. The BI is still read and its result followed up.
Question 15
- Let it sit on the cart until it is dry, then release it
- Wipe the outside of the wrap dry and send it to storage
- Release it if the internal chemical indicator changed color
- Consider it contaminated and send it back to be reprocessed
Show answer
Answer: D. Consider it contaminated and send it back to be reprocessed
Moisture can carry microorganisms through the wrap, so a wet pack is considered non-sterile. The tray is reprocessed, and the cause (load configuration, set weight, steam quality, drying time or cooling conditions) is investigated and documented.
Question 16
- Remove the sterilizer from service and notify the supervisor right away
- Assume it is a false positive because the printout and indicators passed
- Keep using the sterilizer and recheck it with another BI next week
- Discard the result and record only a repeat BI that reads negative
Show answer
Answer: A. Remove the sterilizer from service and notify the supervisor right away
A positive BI means the cycle may not have killed all the spores. The sterilizer is taken out of service, and the supervisor and infection prevention are notified. Affected loads are identified and quarantined or recalled per policy. The cause is investigated, the sterilizer retested, and everything documented.
Test yourself under exam conditions