CRCST Preparation and Packaging practice questions
Inspecting instruments for cleanliness and function, assembling sets with count sheets and indicators, choosing and closing packaging, and labeling packages correctly. This domain is about 21% of the exam. Try each question, then open the answer.
Read the Preparation and Packaging study guide
Question 1
- Repackage them and have the sealer's settings checked
- Cover the wrinkled part of each seal with a strip of indicator tape
- Run each pouch through the sealer again over the same seal
- Sterilize the pouches and recheck the seals after the cycle
Show answer
Answer: A. Repackage them and have the sealer's settings checked
A good heat seal is continuous and uniform, with no wrinkles, channels, or gaps, because any channel can let microorganisms in after sterilization. Defective pouches are repackaged. Repeated defects suggest a sealer problem, so the settings should be checked against the manufacturer's IFU.
Question 2
- To confirm the grasper's jaws close with even tension
- To find insulation breaks that could burn the patient
- To verify that the instrument's lumen is free of soil
- To check that the instrument is dry enough to package
Show answer
Answer: B. To find insulation breaks that could burn the patient
Insulated electrosurgical instruments can have tiny cracks or pinholes in their coating that are hard to see. Current can escape through these breaks and burn tissue away from the surgical site. An insulation tester finds breaks that visual inspection can miss.
Question 3
- Spray it with instrument lubricant to loosen the debris
- Package it, since steam will penetrate the dried debris
- Wipe the box lock with a damp cloth at the prep station
- Return it to decontamination for recleaning
Show answer
Answer: D. Return it to decontamination for recleaning
Inspection in the prep area is a check on cleaning. Any instrument with visible soil goes back to decontamination to be recleaned, because you cannot sterilize what is not clean. Box locks, serrations, and hinges are common places for soil to hide, which is why they are inspected under magnification.
Question 4
- They are too thick to reach inside box locks
- Oils and silicone leave a film that blocks sterilant
- Mineral oil turns dark when heated and stains the tray
- Silicone dissolves stainless steel's passivation layer
Show answer
Answer: B. Oils and silicone leave a film that blocks sterilant
Mineral oil and silicone leave a coating that steam and other sterilants cannot pass through, so microorganisms under the film may survive. Lubricants are chosen and used per the instrument and lubricant IFUs.
Question 5
- 300-series (austenitic) stainless steel
- 400-series (martensitic) stainless steel
- Anodized aluminum alloy
- Chrome-plated brass
Show answer
Answer: B. 400-series (martensitic) stainless steel
400-series (martensitic) stainless steel can be heat-hardened, so it keeps a cutting edge and holds up to clamping stress. 300-series (austenitic) stainless steel is more corrosion resistant but softer, so it is used for retractors, basins, and other non-cutting items. Knowing this helps explain why some instruments need more care.
Question 6
- Leave all seven, since an extra instrument does no harm
- Change the count sheet to show seven hemostats
- Discard the extra hemostat to keep the count correct
- Remove the extra one and route it per policy
Show answer
Answer: D. Remove the extra one and route it per policy
Sets are assembled to match the count sheet exactly, because the surgical team relies on accurate counts. An extra instrument usually belongs to another set, so it is routed per policy to where it belongs. Count sheets are changed only through the approved process.
Question 7
- There is no problem, because tight guards cannot fall off
- Tip protectors are allowed only in low-temperature sterilization
- They may block steam; use validated, permeable protectors
- Tip guards should be added after sterilization, not before
Show answer
Answer: C. They may block steam; use validated, permeable protectors
Tip protectors guard sharp or delicate tips. They must be made of material the sterilant can pass through, be validated for the sterilization method, and fit without sealing off the tips. Tight, non-vented guards can keep steam from contacting the surfaces under them.
Question 8
- Nest same-size basins tightly together so that no air gaps remain between them
- Alternate the direction each basin faces so the weight is balanced in the set
- Place the heaviest instruments inside the top basin to make good use of space
- Separate with absorbent material; 1-inch size difference; all facing one way
Show answer
Answer: D. Separate with absorbent material; 1-inch size difference; all facing one way
Nested basins need space between them so steam can reach every surface and condensate can drain. A difference of at least 1 inch in diameter, absorbent material between basins, and all basins facing the same way make this possible.
Question 9
- That the pack was exposed to the process
- That the contents of the pack are sterile
- That all critical variables were met inside the pack
- That the sterilizer passed its air-removal test
Show answer
Answer: A. That the pack was exposed to the process
External indicator tape is a Type 1 process indicator. Its color change shows that the package was exposed to the sterilization process, which separates processed packs from unprocessed ones. It does not show that the contents are sterile.
Question 10
- On top of the upper basket only, where the user sees it first
- Taped to the outside of the container lid
- In the bottom level only, since it heats last
- In each level, at the hardest-to-reach spot per IFU
Show answer
Answer: D. In each level, at the hardest-to-reach spot per IFU
An internal chemical indicator goes where the sterilant is least likely to reach, as stated in the IFU. That is often the geometric center, or opposite corners in containers. In multi-level sets, each level gets its own indicator, because conditions can differ between levels. The user then checks each indicator when the set is opened.
Question 11
- 15 lb (6.8 kg)
- 30 lb (13.6 kg)
- 25 lb (11.3 kg)
- 40 lb (18.1 kg)
Show answer
Answer: C. 25 lb (11.3 kg)
AAMI ST79 and ST77 limit instrument sets to 25 lb (about 11.3 kg), counting the container or wrap. Heavier sets are harder to sterilize and dry and are a lifting hazard for staff. Sets over the limit are split or reconfigured.
Question 12
- Reuse it once, since it never contacted a patient
- Reuse it if it passes a careful visual check
- Launder it with the woven textiles, then reuse it
- Discard it; disposable wrap is made for one use
Show answer
Answer: D. Discard it; disposable wrap is made for one use
Disposable nonwoven wraps are made and cleared for a single use. Handling, folding and the sterilization cycle can weaken the material and leave damage too small to see. Nothing validates the wrap for a second cycle. Only reusable woven textiles are laundered and inspected between uses.
Question 13
- Wrapping with one double-layer wrap folded as a single sheet
- Wrapping several sets one after another in the same wrap
- Wrapping with two separate wraps, one after the other
- Placing a peel pouch inside a single wrap
Show answer
Answer: C. Wrapping with two separate wraps, one after the other
Sequential wrapping uses two separate wraps. The item is wrapped completely in the first wrap, then wrapped again in the second, making a package within a package.
Question 14
- Two pouches of the same size, with the inner pouch folded to fit
- A smaller inner pouch, not folded, placed paper side to paper side
- A smaller inner pouch placed plastic side facing the outer paper side
- A larger inner pouch with its seal folded under the item
Show answer
Answer: B. A smaller inner pouch, not folded, placed paper side to paper side
In double pouching, the inner pouch is smaller so it fits without folding. The paper sides face the same way (paper to paper) so the sterilant can pass through both layers and air can be removed. The contents can then be seen through both plastic sides.
Question 15
- Use it, since the filter is the only barrier that matters
- Wrap the container in one layer of wrap and use it
- Seal the crack with indicator tape and use it
- Take it out of use until the gasket is replaced
Show answer
Answer: D. Take it out of use until the gasket is replaced
Rigid containers are inspected before each use: gaskets, filters or valves, latches, and locks. A damaged gasket means the lid will not seal, so the container cannot keep its contents sterile. It is removed from service and repaired per the container IFU.
Question 16
- Directly on the wrap with a ballpoint pen
- On the wrap with a pencil so it can be erased
- On a sticky note tucked under the final fold of the wrap
- On indicator tape or a label, with a felt-tip marker
Show answer
Answer: D. On indicator tape or a label, with a felt-tip marker
Label information goes on the indicator tape or on a label made for sterilization, never directly on the wrap. A felt-tip, non-toxic, indelible marker is used so the writing lasts and does not damage the package. Sharp pens can puncture packaging.
Test yourself under exam conditions