CRCST Cleaning, Decontamination and Disinfection practice questions
Bloodborne pathogen safety, microbiology and PPE, manual and mechanical cleaning, water quality, cleaning-equipment testing and maintenance, and the Spaulding levels of disinfection. This domain is about 21% of the exam. Try each question, then open the answer.
Read the Cleaning, Decontamination and Disinfection study guide
Question 1
- Have the employee pay and then submit the cost for reimbursement
- Require the vaccine with no option for the employee to decline
- Offer the vaccine only after an exposure incident has occurred
- Offer the vaccine series at no cost to the employee
Show answer
Answer: D. Offer the vaccine series at no cost to the employee
The Bloodborne Pathogens Standard requires employers to offer the hepatitis B vaccine series to workers with occupational exposure, at no cost to the worker. An employee who declines signs a declination statement and can still accept the vaccine later at no cost.
Question 2
- Susceptible host
- Reservoir
- Portal of exit
- Mode of transmission
Show answer
Answer: A. Susceptible host
The chain of infection has six links: infectious agent, reservoir, portal of exit, mode of transmission, portal of entry, and susceptible host. Vaccination protects the worker, so the worker is no longer a susceptible host for that virus. Breaking any one link stops the spread of infection.
Question 3
- The cell walls of gram-negative bacteria
- The protective coat of bacterial spores
- The outer envelope of viruses such as HIV
- The cell walls of fungi such as Candida
Show answer
Answer: A. The cell walls of gram-negative bacteria
Endotoxin is a lipopolysaccharide in the outer membrane of gram-negative bacteria. It is released when the bacteria die and it is heat-stable, so ordinary sterilization does not destroy it. Prompt cleaning and good water quality keep bacterial loads, and therefore endotoxin, low before sterilization.
Question 4
- Cloth lab coat, exam gloves tucked under the coat cuffs, surgical mask, hair cover, and shoe covers
- Fluid-resistant gown, heavy-duty gloves, hair cover, shoe covers, with eye protection only at the ultrasonic
- Scrub top, heavy-duty gloves, safety glasses without side shields, hair cover
- Fluid-resistant long-sleeved gown, heavy-duty gloves over gown cuffs, mask with face shield, hair and shoe covers
Show answer
Answer: D. Fluid-resistant long-sleeved gown, heavy-duty gloves over gown cuffs, mask with face shield, hair and shoe covers
Manual cleaning creates splashes and aerosols, so decontamination attire must protect skin, eyes, nose, and mouth. That means a fluid-resistant gown with sleeves and heavy-duty gloves whose cuffs extend over the gown cuffs. Add a fluid-resistant mask with eye protection (goggles or a full-length face shield). Finish with a hair cover, and shoe covers or dedicated shoes. The employer provides this PPE.
Question 5
- About 1 minute
- At least 15 minutes
- About 5 minutes
- Until the stinging stops
Show answer
Answer: B. At least 15 minutes
Eyewash stations meeting ANSI/ISEA Z358.1 deliver a flushing flow for at least 15 minutes, the usual minimum for a chemical splash. The product's SDS first-aid section may call for longer. Plumbed units are activated weekly to confirm they work.
Question 6
- Faster protein breakdown, so the soak time can be cut in half
- A solution strong enough to also disinfect the instruments
- No change at all, because enzymes cannot be overconcentrated
- More residue to rinse off, with no proven gain in cleaning
Show answer
Answer: D. More residue to rinse off, with no proven gain in cleaning
Detergents are tested and labeled at a specific dilution, water temperature, and contact time. Using more than the label calls for does not improve cleaning and leaves more residue to rinse off. Residues can harm devices or patients.
Question 7
- Treat it as not disinfected, correct the cause, and reprocess it
- Release it, since the cycle ran all the way to completion
- Wipe the outside with alcohol and release it
- Release it and record the alarm in the AER log
Show answer
Answer: A. Treat it as not disinfected, correct the cause, and reprocess it
A channel-flow alarm means solution may not have reached every channel, so high-level disinfection cannot be assumed. The technician checks the connectors and channels for blockage or misconnection, then reprocesses the scope per its IFU. The alarm and the corrective action are documented.
Question 8
- They spin freely and their nozzles are not clogged
- They are oiled with instrument lubricant before the first cycle
- They are locked in place for even coverage
- They are removed so trays fit more easily
Show answer
Answer: A. They spin freely and their nozzles are not clogged
Spray arms deliver water and detergent to every part of the load, so they must turn freely and have open nozzles. Debris, mineral buildup, or an item sticking up from a tray can block or stop an arm. Checking them daily per the washer's IFU catches problems before loads come out dirty.
Question 9
- Close ratchets on the first tooth to keep instruments in place
- Stack trays closely so more fit in each cycle
- Cover the rack with a towel to hold small items in place
- Open hinged instruments and leave room for the spray
Show answer
Answer: D. Open hinged instruments and leave room for the spray
Washer-disinfectors clean by spraying water and detergent onto every surface. Hinged instruments are opened so box locks and jaws are exposed, and racks are not overloaded so the spray reaches everything. Items are placed as the washer and device IFUs direct.
Question 10
- A stainless steel needle holder with serrated jaws
- A set of stainless steel Army-Navy retractors
- A stainless steel hemostat after manual pre-cleaning
- A rigid telescope with optical lenses
Show answer
Answer: D. A rigid telescope with optical lenses
Ultrasonic cleaning is meant for stainless steel instruments with crevices, hinges and serrations. The vibration can damage the lenses of rigid telescopes and the seals around them, so most telescope IFUs do not allow it. Check each device's IFU before placing it in the ultrasonic.
Question 11
- Endophthalmitis caused by a sterilization failure
- Creutzfeldt-Jakob disease spread by instruments
- A latex allergy to the surgeon's gloves
- Toxic anterior segment syndrome (TASS)
Show answer
Answer: D. Toxic anterior segment syndrome (TASS)
TASS is an acute, sterile inflammation of the front (anterior segment) of the eye that usually appears within a day or two of surgery. It is a toxic reaction to something that entered the eye during surgery, not an infection, so antibiotics do not prevent it.
Question 12
- Process them routinely, since standard steam cycles destroy prions
- Use immediate-use steam sterilization so they are processed quickly
- Follow the facility's prion policy and keep them moist
- Let them dry before transport to reduce the aerosol risk
Show answer
Answer: C. Follow the facility's prion policy and keep them moist
Prions resist routine processing, so facilities keep a written CJD/prion policy developed with infection prevention. It often calls for single-use items where possible and special cleaning and sterilization protocols for high-risk tissue. Instruments are kept moist and not allowed to dry, since dried prion-contaminated soil is very hard to remove. SPD should be told about the case in advance so it can prepare.
Question 13
- After high-level disinfection, just before it goes into storage
- Only when the scope shows visible damage
- After manual cleaning, just before it goes into the AER
- After pre-cleaning and before immersion in cleaning solution
Show answer
Answer: D. After pre-cleaning and before immersion in cleaning solution
A leak test checks that the scope's outer covering and internal channels are intact. It is done after every use, after pre-cleaning and before the scope is immersed in cleaning solution. That way a damaged scope does not take in fluid.
Question 14
- Sterilization, because it enters the body
- Intermediate-level disinfection
- High-level disinfection
- Low-level disinfection
Show answer
Answer: C. High-level disinfection
Under the Spaulding classification, items that touch mucous membranes or non-intact skin are semi-critical and need at least high-level disinfection after cleaning. Sterilization is acceptable for them but goes beyond the minimum.
Question 15
- Once a week, on a set day
- Only when the solution looks cloudy
- Only on the day the solution is first poured
- Before each use, as the product IFU directs
Show answer
Answer: D. Before each use, as the product IFU directs
Reusable high-level disinfectants are diluted by water carried in on devices and lose strength over time. MEC test strips confirm the solution is still at or above its minimum effective concentration. They are used before each use per the IFU. Results are documented each time.
Question 16
- Add fresh concentrate to the basin to bring it back up to full strength
- Use it, but double the immersion time to compensate
- Discard it, refill with fresh solution, retest, and document
- Use it for today's scopes and change it tomorrow
Show answer
Answer: C. Discard it, refill with fresh solution, retest, and document
A failed MEC test means the solution can no longer be relied on to achieve high-level disinfection. It is discarded, replaced with fresh solution prepared per IFU, retested, and the failure and corrective action are documented. No devices are processed in the failed solution.